Lux and luminance meter calibration
LuxCal250 from BenthamThe LuxCal250 provides accurate luxmeter calibration over an exceptionally wide dynamic range. Comprising a light-tight enclosure housing standards of illuminance, Bentham’s LuxCal250 presents a straightforward and cost-effective calibration solution obviating requirements typical of traditional, distance-based calibration.Designed as a high-performance, all-in-one calibration solution, the capabilities of the LuxCal250 may be extended to encompass precise luminance meter calibration with the LumiCal50, alongside representing a UKAS accredited calibration device with the LuxCal+.
- Houses standards in light tight enclosure with easy access to position devices under test
- Colour shifted (2856K quartz tungsten halogen (QTH) standards (CIE illuminant A)
- Output level of the illuminance / luminance calibrator is varied via neutral density filters,
- Lab jack and height gauge, featuring a laser alignment aid, allow accurate positioning of device under test
- Calibration traceable to Physicalish Technische Bundesanstalt (PTB, Germany)
Further information
Core benefits
- Calibrate luxmeter/ luminance meters in accordance with ISO/CIE 19476:2014
- No darkroom or optical bench required to achieve wide dynamic range
- Calibration traceable to PTB, Germany
- Optional luminance or UKAS accredited calibration
Reported parameters
- Illuminance (lx)
- Optional Luminance (cd.m-2)
Specifications
LuxCal250
Lamp type | Quartz Tungsten Halogen lamp, G6.35 base |
Nominal lamp power and voltage | 250W, 12V |
Operating current | 10.4A (adjusted to obtain 2856K operation) |
CCT | 2856±10K |
Expected lifetime | 2000 hours |
Calibration frequency | 100 hours use/ one year recommended |
Dimensions | 600L x 500W x 700H (mm) |
Typical illuminance level one (no filter) | 19000 lx |
Typical illuminance level two (ND1) | 1800 lx |
Typical illuminance level three (ND2) | 180 lx |
Typical illuminance level four (ND3) | 20 lx |
Typical illuminance level five (ND4) | 2 lx |
LumiCal50
Lamp type | Grit-blasted Quartz Tungsten Halogen lamp, G6.35 base |
Nominal lamp power and voltage | 50W, 12V |
Operating current | 4.0A (adjusted to obtain 2856K operation) |
Expected lifetime | 2000 hours |
Calibration frequency | 100 hours use/ one year recommended |
Typical luminance level one (no filter) | 21000 cd.m-2 |
Typical luminance level two (ND1) | 1800 cd.m-2 |
Typical luminance level three (ND2) | 200 cd.m-2 |
Typical luminance level four (ND3) | 30 cd.m-2 |
Typical luminance level five (ND4) | 3 cd.m-2 |
LuxMeter+
Spectral function | V(λ) photometric |
V(λ) matching error | f1 ≤ 3 % |
Typical responsivity | 0.5 nA/lx |
Diffuser diameter | 7mm |
Cosine matching error | f2 ≤ 1.5 % |
Typical minimum detectable illuminance | 20 mlx |
Typical maximum detectable illuminance | 20 klx |
Display | 5 digits, auto-ranging |
Power supply | Mains transformer |
Interface | USB 2.0 |
610 Constant current power supply
Input voltage | 100V - 220V, 5A max, 50-60Hz |
Output voltage & current | 0-10.4A in remote use. Max Voltage: 26V |
Connection type | IEC mains input connection |
Resolution | 0.1A local; 0.001A over SCPI protocol |
Operation modes | Remote mode |
Current pre-set | 4.00 A, 5.40 A, 6.30 A, 8.50 A, 10.40 A, 0-10.4A variable |
Max output power | 250W |
Setting control | Remote setting by user interface controlled through USB |
Operating conditions | Ambient temperature: 0°C - +40°C |
Drift | 0.05% of rated output |
Temperature stability | 30ppm/°C |
Compliance to standards | EN 61010-1:2010 |
Accuracy | ±(0.04% +0.05A) |
Output power ripple | <0.1% |
Dimensions | 450L x 300W x 130H (mm) |
Weight | 5 kg |
Mode types | Constant Current, Constant Voltage |
Applications
- ISO/CIE 19476:2014, Characterization of the performance of illuminance meters and luminance meters
- CIE S 023/E:2013, Characterization of the performance of illuminance meters and luminance meters
- Luxmeter calibration
- Luminance meter calibration
- UKAS accredited calibration