In Vitro sunscreen spectrophotometer

SSUV300 from Bentham

The SSUV300 boasts numerous advantages in the performance of accurate measurement of UV absorbance of thin film sunscreen samples. Including a dual monochromator-based spectrophotometer, a simple magnetic PMMA plate carrier and integrating sphere, collection of transmitted light and accurate measurements in accordance with international standards is ensured. The challenging measurement of sunscreen products, including high absorbance, fluorescing samples and diffuse transmission properties are fully taken account of in the design of the SSUV300.

Features
  • Split double monochromator configuration
  • Monochromatic sample illumination (290-400nm)
  • Integrating sphere collection of transmitted light
  • Monochromator detection (290-400nm)
  • Fully automated and reporting of results

Further information

Core benefits

  • Measure absorbance down to 6 OD
  • Avoid incorrect results due to fluorescence
  • Easy to use
  • Accurate and repeatable measurements

Reported parameters

  • UV absorbance
  • UVAPF
  • Critical wavelength 
  • UVA/UVB ratio

Specifications

Measurement

Spectral range of operation

290-400nm

Bandwidth

1nm

Wavelength accuracy

± 0.1nm (2400g/mm)

Beam diameter at sample plane

4mm diameter

Measured absorbance range

0-6 OD

PMMA plate dimensions

5x5cm

Bench space required

1m deep x 2.5m wide

Services requirements

5 x main sockets 1200W power consumption

Monochromatic probe source

UV source

150W short-arc xenon lamp

Monochromator configuration

Symmetric, single Czerny-Turner

Monochromator focal length

300mm

Diffraction gratings

2400 g/mm

Dispersion

5.4nm/mm (2400g/mm)

Wavelength accuracy

± 0.1nm (2400g/mm)

Beam diameter at sample plan

4mm diameter

Computer requirements

OS

Windows 7, Windows 8 (32-/64-bit) operating systems (or newer)

Software control

BenWin+ Spectral acquisition software

Minimum hard disk space

Approx. 100MB

Minimum RAM

2 GB

Interface

4 x USB 2.0 ports

Applications

  • ISO 24443:2012 - Determination of sunscreen UVA photoprotection in vitro
  • EN ISO 24443:2012 - Determination of sunscreen UVA photoprotection in vitro
  • AS/NZS 2604:2012 - Sunscreen products - Evaluation and classification
  • FDA, Final Rule 2011 - Rulemaking History for OTC Sunscreen Drug Products

Downloads

B-SSUV300 In vitro sunscreen spectrophotometer

Contact

Ing. Fabrizio Renzi
Ing. Fabrizio Renzi

Contact

Quantum Design s.r.l.

Via Francesco Sapori, 27
00143 Roma
Italy

Phone:+39 06 5004204
Fax:+39 06 5010389
E-mail:italy@qd-europe.com
Ing. Fabrizio RenziProduct Manager
+39 06 5004204
Write e-mail