Photometer characterisation solution

PCS300 from Bentham

The PCS300 provides accurate characterisation of key parameters relating to the performance of luxmeters, luminance meters and components.

Set around a tuneable monochromatic light source, the PCS300 enables the spectral responsivity measurement of photometers to determine their deviation from the desired CIE V(λ) function, represented by the f1’ mismatch index. Furthermore, product development is empowered by the possibility of measuring the component detectors and filters of photometers, ensuring blocking out of band. The addition of a motorised goniometer allows the measurement of directional response.

Features
  • Xenon lamp/ single monochromator tuneable light source
  • Dual optical path optic to couple light to DUT and reference photodiode for parallel measurement (luxmeters)
  • Integrating sphere accessory for characterisation of luminance meter spectral mismatch
  • Collimated deuterium/ halogen source option for measurement of filter transmission
  • Motorised single axis goniometer with manual DUT 90° rotation for directional response measurement
  • Calibration traceable to Physicalish Technische Bundesanstalt (PTB, Germany

Further information

Easy to use, the PCS300 will directly report the values important to demonstrate the performance of your photometers. Compliment the PCS300 with the LuxCal250 for the calibration of lux and luminance meters according to ISO/CIE 19476:2014.

Core benefits

  • Characterise key parameters of photometers in accordance with ISO/CIE 19476:2014
  • Easy to use and interface devices under test
  • No darkroom or optical bench required (except for directional response measurement)
  • Calibration traceable to PTB, Germany

Reported parameters

  • General V(λ) mismatch index, f1
  • Directional response index, f2
  • filter transmission (%)
  • detector responsivity (A.W-1)

Specifications

Tuneable light source

Probe light source

150W xenon

Monochromator configuration

Triple grating, symmetric, single Czerny-Turner, 300mm focal length

Bandwidth

Adjustable fixed slit, 1-10nm typical

Wavelength range

2400 g/mm grating: 250-320nm
1800 g/mm grating: 320-800nm
1200 g/mm grating: 800-1100nm

Resolution

250-320nm: 0.15nm
320-800nm: 0.2nm
800-1100nm: 0.3nm

Dispersion

250-320nm: 1.35nm/mm
320-800nm: 1.8nm/mm
800-1100nm: 2.7nm/mm

Wavelength accuracy

250-320nm: ± 0.1nm
20-800nm: ± 0.15nm
800-1100nm: ± 0.2nm

Picoammeter

Channel 1 input

Current input to picoammeter

Channel 2 input

Current input to picoammeter

Gain ranges

1010-105 V/A

Maximum current Input

100µA

Frequency response

DC to 30Hz

Gain accuracy

+1%

Gain stability

200ppm/°C

Output stability

5ppm/°C to 500ppm/°C depending on sensitivity

Linearity

< 0.025% departure from linearity from zero to full scale

ADC resolution

4½ digit BCD (0 to 19999) i.e. > 14 bit resolution

ADC integration time

100ms

Picoammeter input impedance

Virtual earth

PCS300_0001 luxmeter measurement accessory

Probe beam

Size 3x3mm

Beam splitter

Synthetic fused silica window

Reference photodiode

Si-CAL spectral responsivity standard

Calibration type and range

Short-circuit spectral responsivity, 360-830nm

Calibration traceability

Physicalish Technische Bundesanstalt (PTB, Germany)

DUT mount

Custom mount.

Software report

Spectral responsivity, f1

PCS300_0002 luminance meter measurement accessory

Sphere diameter

100mm

Sphere exit port diameter

10mm

Sphere coating

BaSO4

Calibration type and range

Short-circuit spectral responsivity, 360-830nm

Calibration traceability

Physicalish Technische Bundesanstalt (PTB, Germany)

Software report

Spectral responsivity, f1

PCS300_0003 directional response accessory

Source

CL6-H enclosed 150W Halogen lamp

Typical lamp distance

2m

DUT mount

Custom mount.

DUT rotation

One axis motorised, manual rotation of mount through 90°

Angular accuracy

≤0.1°

Angular resolution

≤0.001°

Software report

Directional responsivity, f2

PCS300_0004 filter transmission accessory

Illumination type

Polychromatic, collimated beam

Light source

ILD-D2-QH deuterium-halogen source

Collimator

COL_2_UV-VIS collimator

Detectors

DH_30 multi-alkali photomultiplier & DH_Si silicon photodiode

Sample holder

Custom sample holder up to 15mm diameter.

Applications

ISO/CIE 19476:2014, Characterization of the performance of illuminance meters and luminance meters

Contact

Ing. Fabrizio Renzi
Ing. Fabrizio Renzi

Related products

Lux and luminance meter calibration
The LuxCal250 provides accurate luxmeter calibration over an exceptionally wide dynamic range. Comprising a light-tight enclosure housing standards of illuminance, Bentham’s LuxCal250 presents a ...

Contact

Quantum Design s.r.l.

Via Francesco Sapori, 27
00143 Roma
Italy

Phone:+39 06 5004204
Fax:+39 06 5010389
E-mail:italy@qd-europe.com
Ing. Fabrizio RenziProduct Manager
+39 06 5004204
Write e-mail