Good to know about KBr pellets

KBr pellet formation is a classic technique for the analysis of solids that retains attractive features over newer techniques such as ATR. The ability to change the pathlength of the compound of interest offers a substantial advantage.

Advantages:

  • Use less sample than ATR
  • Higher signal-to-noise
  • Control signal intensity by changing sample concentration or by increasing the pathlength by adding extra sample and KBr to the pellet die. The intensity increases as the mass increases. In accordance with the Beer-Lambert law absorbance increases linearly with the pellet mass which is proportional to pathlength. This gives the operator control over the peak intensities. When attempting to identify weak bands, possibly from trace contaminants, KBr pellets offer clear advantages with substantial improvements in limits of detection.
  • No peak intensity correction required (in ATR spectroscopy the effectively pathlenth is dependent on the wavelength, resulting in changes of the relative band intensities and, for wide bands, in a shift of the absorption maximum)
  • A variety of (commercial) spectra libraries

Further information

Preparation of pellets

Generally, it is easy to produce a good quality KBr pellet if an evacuable pellet die is used correctly. When the sample is added to the halide salt the clarity of the disc will depend to a large extent on the quantity and type of sample. For a 13mm diameter die 200 - 300 mg KBr and 1 mg sample at load of 8 tons are a good starting point. To ensure that a sample pellet is produced which will enable accurate spectra of samples to be obtained, it is essential that the sample is thoroughly blended with the halide salt powder. Blending can be achieved either by using a mortar and pestle or by using a grinding mill.

The overall quality of a pellet is largely dependent upon the quality of the KBr or halide salt powder used, which should always be of a spectroscopic grade of purity.

Good to know – Possible faults and remedy

However, some faults in the produced sample pellet may occur due to a variety of reasons. Some of these faults and their remedies are tabulated below. The faults described are for pure KBr or other halide salts which do not contaminate the sample.

Fault

Cause

Remedy

Sample pellet not clear. Lacks optical clarity.

Sample damp, contaminated KBr powder or insufficient
pressure when compacting.

Dry the KBr powder or sample and increase the compacting
pressure.

Sample pellet is clear but shows opaque spots.

Powder not uniformly flat in the die, leaving large particles which do not vitrify when
pressed.

Sieve powder to extract coarse grains, then re-grind and re-press.

Sample pellet is cloudy.

Insufficient evacuation time or leaky seals.

Check seals on the die and lengthen evacuation period.

Sample pellet is clear at first but quickly becomes cloudy.

Damp powder or damp atmosphere.

Dry the KBr powder or sample,
check seals on the die and lengthen evacuation period.

Contact

Dr. Joachim Weiss
Dr. Joachim Weiss

Related products

Press dies for producing KBr pellets and for x-ray spectroscopy sample preparation
We provide evacuable pellet press dies made of hardened stainless steel for the preparation of high-quality pellets with diameters between 5 mm - 40 mm.  The press dies come with all necessary ...
Laboratory presses with 15- and 25- ton loads
Our robust manual hydraulic laboratory presses with 15 and 25 t load configurations are essential for producing KBr pellets and the preparation of XRF samples. They have been in the market for decades ...
8 t/ 15 t/ 25 t power assisted hydraulic press
Atlas is a power assisted hydraulic press and is available in three version for compressive loads up to 8 t, 15 t and 25 t. The press is optimized for IR and XRF applications and can be used with all ...
Mini lab press
The Mini Lab Press is an innovative and low-cost instrument for producing 7 mm KBr pellets. The press is designed as a table-top unit and operates hydraulically. The desired compressive load can be ...

Contact

Quantum Design GmbH

Breitwieserweg 9
64319 Pfungstadt
Germany

Phone:+49 6157 80710-0
Fax:+49 6157 807109
E-mail:germanyqd-europe.com
Dr. Joachim WeissProduct Manager - Spectroscopy & Marketing Manager - Content
+49 6157 80710-72
Write e-mail