Night vision compatibility testing
NVC150 from BenthamThe NVC150 is an essential tool in the evaluation and development of night vision enabled products, ranging from goggles through to displays.
Featuring two configurations, as a spectroradiometer, the NVC150 permits accurate assessment of interior lighting (MIL STD 3009) and exterior lighting (SAE ARP5825, STANAG 1445 etc) whilst as a spectrophotometer, is utilised in the detailed evaluation of NVIS filter transmission.
- Double monochromator for superlative stray light performance
- Thermoelectrically cooled photomultiplier detector (S20 photocathode)
- Optional spectroradiometer and spectrophotometer modes
- Range of telescope input optics
- Reports colourimetric parameters, luminance, NRa and NRb
Further information
Across all applications, the superlative performance of the NVC150, including excellent stray light rejection and unparalleled sensitivity, delivers accurate NVIS compatibility characterisation and ultimate measurement confidence across the life time of the system.
Core benefits
- Easy to use Windows control
- Ultimate measurement confidence
- Complies with legislative requirements
- Assess all types of lighting product (interior & exterior)
Reported parameters
- Colourimetric parameters
- Luminance
- NRa
- NRb
Specifications
Spectral range of operation | 380-930nm |
Spectral data interval | 5nm |
Spectral bandwidth (FWHM) | 5nm |
Wavelength accuracy | ± 0.3nm, ± 0.05nm with software correction |
Wavelength reproducibility | ± 0.05nm |
Resolution | 0.5nm |
Stray light rejection at 2.5 FWHM | 10-8 |
Minimum working distance | 25mm with Macro lens |
Maximum working distance | 50m with TL1 lens |
Typical NER | 10-10 W.m-2.sr-1 |
Bench space required | 1m deep x 2.5m wide typical including working space |
Services requirements | 2 x 110/220V AC mains sockets, 400VA total |
Applications
- MIL STD 3009 - Lighting, aircraft, night vision imaging system (NVIS) compatible
- SAE ARP5825 - Design Requirements and Test Procedures for Dual Mode Exterior Lights
- STANAG 1445 - Conduct Of Maritime Nvg Cross Operations
Downloads
Contact
+49 6157 80710-34 | |
+49 6157 80710934 | |
Write e-mail |
Navigation
Categories
Contact
Quantum Design GmbH
Breitwieserweg 9
64319 Pfungstadt
Germany
Phone: | +49 6157 80710-0 |
Fax: | +49 6157 807109 |
E-mail: | germanyqd-europe.com |